GW Instek PLR-Series Programmable Switching D.C. Power Supply
Model | PLR20-18 | PLR20-36 | PLR36-10 | PLR36-20 | PLR60-6 | PLR60-12 |
OUTPUT RATING | ||||||
Voltage | 0V to 20V | 0V to 36V | 0V to 60V | |||
Current | 0 ~18A | 0 ~36A | 0 ~10A | 0 ~20A | 0 ~6A | 0 ~12A |
Power | 360W | 720W | 360W | 720W | 360W | 720W |
REGULATION (CV) | ||||||
Load | 0.005%FS+2mV | 0.005%FS+2mV | 0.005%FS+2mV | |||
Line | 0.005%FS+1mV | 0.005%FS+1mV | 0.005%FS+1mV | |||
REGULATION (CC) | ||||||
Load | 5mA | 5mA | 5mA | |||
Line | 5mA | 10mA | 1mA | 5mA | 1mA | 5mA |
RIPPLE & NOISE (Noise Bandwidth= 20MHz ; Ripple Bandwidth = 1MHz) | ||||||
CV p-p | 30mVp-p | 30mVp-p | 30mVp-p | |||
CV rms | 0.5mVrms | 0.5mVrms | 0.5mVrms | |||
CC rms | 10mArms | 5mArms | 10mArms | 5mArms | 5mArms | |
SETTING ACCURACY (23°C±5°C, after 30-minute aging) | ||||||
Voltage | ± (0.5%SET+0.5%F.S): | ± (0.5%SET+0.5%F.S): | ± (0.5%SET+0.5%F.S): | |||
Current | ± (1%SET+1%F.S): | ± (1%SET+1%F.S): | ± (1%SET+1%F.S): | |||
Power | ±(0.7%rdg+1.5%F.S): | ±(0.7%rdg+1.5%F.S): | ±(0.7%rdg+1.5%F.S): | |||
READBACK ACCURACY (23°C±5°C, after 30-minute aging) | ||||||
Voltage | ±(0.1%rdg+2digit): | ±(0.1%rdg+2digit): | ±(0.1%rdg+2digit): | |||
Current | ±(0.5%rdg+2digit): | ±(0.5%rdg+2digit): | ±(0.5%rdg+2digit): | |||
RESPONSE TIME | ||||||
Raise Time | 50ms/50ms: No load/Rated load | 50ms/50ms: No load/Rated load | 50ms/50ms: No load/Rated load | |||
Fall Time(Full load) | 50ms ; Output voltage: 90%→10%FS | 50ms ; Output voltage: 90%→10%FS | 150ms ; Output voltage: 90%→10%FS | |||
Fall Time(No load) | 250ms ; Output voltage: 90%→10%FS | 250ms ; Output voltage: 90%→10%FS | 600ms ; Output voltage: 90%→10%FS | |||
Load Transient Recover Time | 100μs | 100μs | 100μs | |||
SETTING RESOLUTION | ||||||
Voltage | 10mV | 10mV | 10mV | |||
Current | 10mA | 10mA | 10mA | |||
MEASUREMENT RESOLUTION | ||||||
Voltage | 10mV | 10mV | 10mV | |||
Current | 10mA | 10mA | 10mA | |||
SERIES AND PARALLEL CAPABILITY | ||||||
Parallel Operation | Up to 3 units | Up to 3 units | Up to 3 units | |||
Series Operation | Up to 2 units | Up to 2 units | Up to 2 units | |||
PPROTECTION FUNCTION | ||||||
OVP | Set range: 10% to 110% FS. Set resolution: 10 times the minimum display resolution. | |||||
OCP | Set range: 5% to 110% FS. Set resolution: 10 times of minimum display resolution. | |||||
UVP | Set range: -1V to 110% FS. Set resolution: 10 times the minimum display resolution. | |||||
ENVIRONMENT CONDITION | ||||||
Operation Temp | 0℃to40℃ | |||||
Storage Temp | ﹣20℃ to 60℃ | |||||
Operating Humidity | 30% to 85% RH (No dew condensation) | |||||
Storage Humidity | 20% to 85% RH (no dew condensation) | |||||
READ BACK TEMP WEFFICIENT | ||||||
Voltage | ±100ppm/℃ | ±100ppm/℃ | ±100ppm/℃ | |||
Current | ±100ppm/℃ | ±100ppm/℃ | ±100ppm/℃ | |||
OTHER | ||||||
Power consumption | 570VA | 1100VA | 520VA | 1050VA | 510VA | 1000VA |
Power factor | 0.99 | 0.99 | 0.99 | |||
Coolingmethod | Forced cooling: Fan speed proportionate to the temperature of the internal heat sink. | |||||
Power Source | Single-phase 100VAC to 240VAC, 50Hz to 60Hz | |||||
Interface | RS-232C/LAN (Standard: ) ; GPIB/USB (option) | |||||
Analog Control | Yes | |||||
Dimension | 139.5 (H) * 140(W) | 139.5 (H) * 210(W) | 139.5 (H) * 140(W) | 139.5 (H) * 210(W) | 139.5 (H) * 140(W) *415.5(D) | 139.5 (H) * 210(W) *415.5(D) |
Weight | 5.2kg | 7.5kg | 5.2kg | 7.5kg | 5.2kg | 7.5kg |